Engineering

The clean room is used for assembly of sensitive instruments and for tests that need to be performed in a clean environment. The thermal vacuum chamber is used for thermal vacuum tests.

 BIRA-IASB clean room

The characteristics of the thermal vacuum chamber and of the clean room:.

BIRA-IASB Grade 1000 clean room
Temperature 18 °C
Humidity < 50%
Cleanliness 100 000

 

BIRA-IASB Thermal vacuum chamber
Useful volume 288 liter, baseplate 60 x 80 cm, height 60 cm
Pumping Rotative (8.3 l/sec) and turbomolecular (1000 l/sec) pump
Attainable pressure 2 10-6 mbar
Flushing Nitrogen flushing possible
Thermal Two independent Hüber control units (for baseplate and shroud)
Thermofluid Silicon oil (-80 °C to +120 °C)
Temperature range -55 °C to +60 °C
Regulation Plate (+/- 0.5 °C), shroud (+/- 2 °C)
Software Siemens driving software

 

The clean room including the thermal vacuum chamber can be hired (incl. operator) by external labs and/or companies.

Prices for use of TVAC and clean room facilities:

  • working days between 9 am and 5 pm (with operator):
    110 euro per hour
  • working days between 5 pm and 9 am (with operator):
    220 euro per hour
  • working days between 5 pm and 9 am (without operator – vacuum and temperature maintained):
    55 euro per hour
  • non working days (Saturday, Sunday, holidays) (with operator):
    220 euro per hour
  • non working days (Saturday, Sunday, holidays) (with operator - vacuum and temperature maintained):
    55 euro per hour

  • (note: the presence of an operator during the night or the week-end is possible only in exceptional cases).

View on the (opened) thermal vacuum chamber with test object inside